Product Name and Code
INCUBATOR FOR BIOLOGICAL TEST
INSTALLATION
•The LISTER biological incubator must be placed on a flat surface.Due to the characteristics of the device, it is important to observethe following measures precautions:
• The device should not be used in places where there is a risk of explosion.
• Remove from the packaging and check the condition of the device. If there are any signs of damage on the device, notity the distributor/manutacturer.
• Place the device on a flat surface, away from sources of heat, i.e. (oven, washer, cooker, etc.)
• Place the device so that the warning light is clearly visible to the user.
• Connect the power cord to a 230 electrical socket and then the 12V cord to a power socket (Figure 1)
• Attention, to power supply use only the power adapter supplied in the setA(12V, (5A). The use of other power supplies may damage themodule and avoid the warranty.
Indications for Use
1. Determine the biological indicator.
2. insert the biological indicator into sterilization in a representative package or as recommended by the autoclave manufacturer.
3. Perform a sterilization cycle.
4. Remove the biological indicator from the autoclave. Allow the indicator to cooldown.
5. Allow the indicator to cool for at least 10 minutes
6. Check for discoloration of the chemical indicator strip (Fig. 2)
7. Smash the glass capsule (Fig. 2) of the biological indicator insidethermoplastic container by squeezing it with pliers or against the edge of the table top. This should be done just before inserting the indicator into theincubator. the indicator should be placed in the incubator no later than one hour after sterilization.
8. insert the biological indicator into the holes in the aluminum block.
9. Start the incubation cycle by pressing the START button, check after a periodof 10 minutes that the LED has changed from intermittent green light to a steadygreen light cycle (otherwise see maintenance and troubleshooting).
10. Incubate the biological indicator for the time described in the indicator manual.
11. For a detalled procedure, please refer to the instructions for use included inside the package of biological indicators.